Original Article

Development and Validation of a Stability Indicating RP-HPLC Method for Simultaneous Estimation of Teneligliptin and Metformin

10.4274/tjps.galenos.2018.16768

  • Rajani VETAPALEM
  • Rajendra Prasad YEJELLA
  • Lakshmana Rao ATMAKURI

Received Date: 17.07.2018 Accepted Date: 06.12.2018 Turk J Pharm Sci 2020;17(2):141-147 PMID: 32454773

Objectives:

The main objective of the present work is to develop a simple, precise, specific and stability method indicating reverse phase high performance liquid chromatography method for simultaneous estimation of teneligliptin and metformin in bulk and tablet dosage form.

Materials and Methods:

The analysis was performed with a Kromasil C18 column (250×4.6 mm, 5 μm) at 30°C using buffer: acetonitrile: methanol (65:25:10, v/v/v) as mobile phase. The detection was carried out with a flow rate of 1.0 mL/min at 254 nm.

Results:

The retention time of teneligliptin and metformin was 2.842 min and 2.017 min, respectively. The linearity range was 5-30 μg/mL for teneligliptin and 125-750 μg/mL for metformin. The forced degradation studies were performed as per the guidelines of the The International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use under acidic, alkaline, oxidative, thermal, photostability, and neutral conditions.

Conclusion:

This method was successfully validated for all the parameters and could detect the the correct amounts of active drug substance in formulations that are available in the market. This developed method in the present study could be successfully employed for the simultaneous estimation of teneligliptin and metformin in bulk and tablet dosage form.

Keywords: Teneligliptin, metformin, RP-HPLC, validation, stability studies

INTRODUCTION

Teneligliptin (TEN) (Figure 1) is chemically [(2S, 4S)-4-[4-(5-methyl-2-phenylpyrazol-3-yl)piperazin-1-yl]pyrrolidin-2-yl]-(1,3-thiazolidin-3-yl) methanone. It is highly effective in lowering blood glucose levels. This drug inhibits the enzyme dipeptidyl peptidase-4, which degrades incretin, a hormone adjusting blood glucose control. It is effectively used to treat type-2 diabetes mellitus.1,2

Metformin (MET) (Figure 2) is chemically 1-carbamimidamido-N,N-dimethylmethanimidamide. It belongs to the biguanide class of antidiabetic drugs. It is the first line drug of choice for the treatment of type-2 diabetes. It activates adenosine monophosphate activated protein kinase, a liver enzyme that plays an important role in insulin signaling, whole body energy balance, and metabolism of glucose and fats.3,4,5

A literature survey reveals a good number of analytical methods for the estimation of TEN and MET individually or in combination with other drugs using ultraviolet (UV) spectrophotometry,6,7,8 high performance liquid chromatography (HPLC),9,10,11,12,13,14,15,16,17,18,19 HPTLC,20 and LC-MS/MS.21 Moreover, methods were reported for the estimation of the selected drugs in their combinations using UV spectrophotometry,22,23 and HPLC.24,25,26,27,28 To the best knowledge of the authors, no stability indicating RP-HPLC method has been reported so far for the simultaneous estimation of TEN and MET. Hence, we tried to develop a simple stability indicating HPLC method for the estimation of the selected drugs. The developed method has been validated as per the guidelines of the ICH.29 To establish the stability indicating nature of the method forced degradation studies were planned for the proposed method under acidic, alkaline, oxidative, thermal, photostability, and neutral conditions.30


MATERIALS AND METHODS


Materials and reagents

Reference standards of TEN and MET were provided as gift samples by Spectrum Labs, (Hyderabad, India). Commercially available tablet formulation Tendia M tablets for the assay studies were purchased from a local pharmacy. HPLC grade methanol, HPLC grade acetonitrile, analytical grade orthophosphoric acid, and HPLC grade water were purchased from Merck Specialities (Mumbai, India). Ethic committee approval was not required for our study.


Instrumentation

The development and validation of the method were performed on a Waters HPLC 2695 system equipped with quaternary pumps, an autosampler, and a photodiode array detector. Empower 2 software was applied for data collection and processing.


Methodology


Statistical analysis

The analytical characteristics of the tested method in HPLC were validated to ensure the suitability of the analytical requirements and reliability of the results. The statistical One Way Variance analysis treatments were performed with the statistical software GraphPad InStat.


Preparation of standard stock solutions

Standard stock solutions of 200 µg/mL for TEN and 5000 µg/mL for MET were prepared by accurately weighing and transferring 2 mg of TEN and 50 mg of MET into 10 mL volumetric flasks. About three fourths of the volume of diluent was added, followed by sonication for 10 min. Finally, the flasks were made up to the mark with diluent to obtain the mentioned concentrations. Next, 1 mL of the above solution was pipetted out and transferred into a 10 mL volumetric flask and diluted up to the mark with diluent to obtain a concentration of 20 µg/mL for TEN and 500 µg/mL for MET.


Preparation of sample solution

Twenty tablets were weighed and average weight was calculated. Then they were powdered using a mortar and pestle and the powder equivalent to 20 mg of TEN and 500 mg of MET was accurately weighed and transferred into a 100 mL volumetric flask. Next 50 mL of diluent was added and the mixture sonicated for 25 min. Further the volume was made up with diluent to obtain a concentration of 200 µg/mL for TEN and 5000 µg/mL for MET. Filters of 0.45 micron size were employed for filtration in the mentioned procedure. Next, 1 mL of the above solution was pipetted out and transferred into a 10 mL volumetric flask and diluted up to the mark with diluents to obtain a concentration of 20 µg/mL for TEN and 500 µg/mL for MET.


Preparation of buffer

One milliliter of orthophosphoric acid was diluted to 100 mL with HPLC grade water to obtain 0.1% orthophosphoric acid buffer.


Mobile phase

Buffer, acetonitrile, and methanol were taken in the ratio of 65:25:10 (v/v/v) and used as mobile phase.


Method validation

Method validation was done as per the guidelines of the ICH.29,30


System suitability

System performance parameters like retention time, number of theoretical plates, tailing factor, and resolution were calculated by injecting standard solutions six times. The resultant results were compared with the standard limits as per the guidelines.


Specificity

Specificity is the ability of a method to discriminate between the analyte of interest and other components that are present in the sample. These studies are performed to check the interferences in the optimized method. To assess the method’s specificity, blank and placebo were injected into the HPLC system under optimized conditions. There should not be any interfering peak in the blank or placebo chromatograms at the retention times of the selected drugs.


Linearity

The linearity of the method was obtained by preparation of the calibration standards of 6 different concentrations in 6 replicates. The calibration curve plots for TEN and MET were obtained by plotting the peak areas on the y-axis and concentrations on the X-axis over the concentration ranges of 5-30 µg/mL for TEN and 125-750 µg/mL for MET. The correlation coefficient should be greater than 0.99.


Accuracy

The accuracy of the method was assessed by recovery experiments by adding a known quantity of pure standard drug to the sample solution and recovering the same in terms of its peak areas. The sample was spiked with standard at levels of 50%, 100%, and 150% of test concentrations. The resultant spiked sample was assayed in triplicate. The % recovery for each level should be 98%-102%.


Precision

Precision is the degree of closeness of agreement between the series of measurements obtained from multiple sampling of the same homogeneous sample under prescribed conditions. It is expressed in terms of standard deviation (SD) or relative SD (RSD). Precision may be a measure of either the degree of repeatability or the reproducibility of the analytical method.


Method precision

Sample solutions were injected under optimized conditions 6 times on 6 different days and their peak areas were recorded. RSD % for the peak areas of the 6 standard injection results should not be greater than 2.


Intermediate precision

Six replicates of sample solutions were injected under optimized conditions on the same day and their peak areas were recorded. RSD % for the peak areas of the 6 replicate injection results should not be greater than 2.


Ruggedness

The ruggedness of the method was determined by carrying out the experiment on different instruments, by different operators, and using different columns of similar types.


Robustness

The robustness of the method was determined by making small deliberate changes in the method like flow rate, mobile phase ratio, and temperature. However, one should not find remarkable changes in the results and the obtained results should be within the ranges in the ICH guidelines.


Effect of variation in flow

A sample was analyzed at 0.9 mL/min and 1.1 mL/min flow rate instead of 1.0 mL/min; the remaining conditions were kept unchanged.


Effect of variation in temperature

Temperature of 25°C and 35°C was maintained instead of 30°C. Samples were injected in triplicate and chromatograms were recorded.


Limit of detection and limit of quantitation (LOD and LOQ)

LOD is the smallest concentration that can be detected but not necessarily be quantified as an exact value. It is calculated using the formula

LOD=3.3 σ/S, where σ=SD; s=slope

LOQ is the lowest amount of analyte in the sample that can be quantitatively determined with precision and accuracy.

LOQ=10 σ/S, where σ=SD; s=slope


Forced degradation studies

TEN and MET standard samples were subjected to degradation under different stress conditions like acidic, alkali, oxidative, thermal, photostability, and neutral conditions.

For acidic and alkali degradation samples were refluxed with 2 N HCl and 2 N NaOH at 60°C for 30 min. For oxidative degradation 20% v/v H2O2 was used and the same was refluxed at 60°C for 30 min. For thermal degradation, a sample was placed in an oven at 105°C for 6 h; for photostability degradation, the drug was exposed to UV light by keeping the sample in a UV chamber for 7 days or 200 W h/m2 in a photostability chamber; for neutral degradation, the drugs were refluxed in water for 6 h at 60°C. All the samples were diluted to obtain a final concentration of 20 µg/mL of TEN and 500 µg/mL of MET. Ten microliters of the samples were injected into the system and the chromatograms were recorded to assess the stability of the sample.


Solution stability

The stability of the drug solution was determined for short-term stability and autosampler stability. Short-term stability was tested by keeping the samples at room temperature (25°C) for 24 h. Autosampler stability was determined by storing the samples for 24 h in the autosampler. Each sample was injected 6 times into the HPLC and the results obtained were compared with the nominal values of QC samples.


RESULTS

The results for the optimized chromatographic conditions are shown in Table 1. The system suitability parameters (tailing factor, retention time, and theoretical plates) were within the acceptance criteria. A summary of the system suitability parameters is given in Table 2. We did not find any interfering peaks at the retention times of TEN or MET (Figure 3), which shows that the method is specific. The quantification was linear in the concentration range of 5-30 µg/mL for TEN with a correlation coefficient of 0.999 (Figure 4) and 125-750 µg/mL for MET with a correlation coefficient of 0.999 (Figure 5). The results for linearity are tabulated in Table 3. The recoveries of TEN and MET were in the range of 99.35-99.94% and 99.80-100.61%, respectively. The results were compared with the guidelines and expressed as percentages and are given in Table 4. The precision of the method is satisfactory as RSD % is NMT 2%. The ruggedness was determined by different analysts and on different days. The results are given in Table 5. No remarkable changes in the results were noted in the robustness studies and hence the method is robust. The results are tabulated in Table 6. The assay results were compared with the labeled claim of TEN and MET marketed formulations and the results are tabulated in Table 7. The LOD and LOQ values were calculated using slope and standard deviation values and the same are tabulated in Table 8.


Forced degradation study

The standard solutions were subjected to different stress conditions as mentioned in the procedure. Under acidic conditions, the drugs showed degradation of about 3.66% for TEN and 3.14% for MET and we noted about 3 degradation peaks (Figure 6). Under alkali conditions, the drugs showed degradation of about 2.75% for TEN and 2.67% for MET and 2 degradation peaks were noted (Figure 7). Under oxidative conditions, the drugs showed degradation of about 1.01% for TEN and 1.62% for MET and 1 degradation peak was noted (Figure 8). Under the remaining conditions, i.e. thermal, photostability, and neutral conditions, the degradation was less than 1% for both drugs and no degradation peak was noted (Figures 9-10 -11). The results of the forced degradation studies are tabulated in Table 9.


Stability studies

The drug solutions were found to be stable for 24 h at 25°C for short-term stability and 24 h for autosampler stability.


DISCUSSION

For method optimization different ratios of acetonitrile and buffer were tried but peak resolution was not achieved. Hence, methanol was used in the mobile phase. Different ratios of orthophosphoric acid buffer, acetonitrile, and methanol were tried, i.e. 65:15:20, v/v/v; 60:20:20, v/v/v; 65:20:15, v/v/v. Finally, it was found that buffer: acetonitrile: methanol in the ratio of 65:25:10, v/v/v, gave good peaks and hence were fixed as the mobile phase. A Kromasil C18 (250×4.6 mm, 5 µm) column, 1 mL/min flow rate, 10 µL injection volume, 30°C column oven temperature, and 254 nm wavelength were fixed as optimized conditions, which were found to be suitable for the separation of peaks. These optimized conditions gave a retention time of 2.842 min and 2.017 min for TEN and MET. All the validation results were as per the limits of the ICH and hence showed the method to be reliable and economical for the estimation of drugs. The effectiveness of the method to separate the degraded peaks from analyte shows its stability indicating nature. The degradation on the lower side, i.e. the degradation percent under all conditions, is in the range of 0.05% to 3.66%, showing the stability of the selected drugs. The RSD % values were less than 2.


CONCLUSION

The method developed possesses all the qualities to be a reliable, rapid, sensitive, specific, and economical method according to the above discussed results and data. The study showed the stability indicating nature of the method with the possible short runtime. Hence, the developed method could be conveniently and effectively used for routine simultaneous estimation of TEN and MET in bulk and pharmaceutical dosage form.


Conflicts of interest: No conflict of interest was declared by the authors. The authors alone are responsible for the content and writing of this article.

  1. Yoshida T, Akhashi F, Sakashita H, Kitajima H, Nakamura M, Sonda S, Takeuchi M, Tanaka Y, Ueda N, Sekiguchi S, Ishige T, Shima K, Nabeno M, Abe Y, Anabuki J, Soejima A, Yoshida K, Takashina Y, Ishii S, Kiuchi S, Fukuda S, Tsutsumiuchi R, Kosaka K, Murozono T, Nakamaru Y, Utsumi H, Masutomi N, Kishida H, Miyaguchi I, Hayashi Y. Discovery and preclinical profile of teneligliptin (3-[(2S,4S)-4-[4-(3-methyl-1-phenyl-1H-pyrazol-5-yl)piperazin-1-yl]pyrrolidin-2-ylcarbonyl]thiazolidine): a highly potent, selective, long-lasting and orally active dipeptidyl peptidase IV inhibitor for the treatment of type 2 diabetes. Bioorg Med Chem. 2012;20:5705-5719.
  2. Kishimoto M. Teneligliptin: a DPP-4 inhibitor for the treatment of type 2 diabetes. Diabetes Metab Syndr Obes. 2013;6:187-195.
  3. Indian Pharmacopoeia, Government of India, Ministry of Health & Family Welfare, Volume-2. Ghaziabad; the Indian Pharmacopoeia Commission; 2007:1358-1359.
  4. Merck. The Merck Index (13th ed). White House Station; Merck & Co Inc; 2001:1061.
  5. Sweetman S, ed. Martindale: The Complete Drug Reference Vol. 1. (36th ed). London; Pharmaceutical Press. 2009:453-454.
  6. Amit MS, Kiran KD, Varsha AR. A simple UV spectrophotometric method development and validation of teneligliptin in tablet dosage form. Indo Am J Pharm Res. 2016;6:14-21.
  7. Manjusha DK, Barhate VD. Spectrophotometric determination of an anti diabetic drug Teneligliptin bulk and pharmaceutical formulations. W J Pharm Res. 2016;5:1625-1632.
  8. Mahadik PS, Senthilkumar GP, Devprakash D, Tamiz MT, Priyanka KG, Sulbha AG. Method development and validation of Metformin in bulk and pharmaceutical dosage forms by using spectrophotometric method. Am J Pharmtech Res. 2012;2:56-61.
  9. Atul TH, Rathod EA, Gupta KR, Umekar MJ. HPLC and UV-spectrophotometric estimation of Teneligliptin from tablet dosage form. J Pharm Biomed Anal. 2016;4:148-156.
  10. Ganeshkumar TNV, Vidyadhara S, Niteen AN, Saisilpa Y, Rajyalakshmi M. Method development, validation, and stability studies of teneligliptin by RP-HPLC and identification of degradation products by UPLC tandem mass spectroscopy. J Anal Sci Tech. 2016;7:18-23.
  11. Chandana M, Prasad Rao M, Samrajyam B, Sireesha KSKD, Naga Premi VV. Analytical method development and validation of teneligliptin in pharmaceutical dosage form by RP-HPLC method. J Heal Sci Nur. 2016;1:1-12.
  12. Atul TH, Rathod EA, Gupta KR, Umekar MJ. HPLC and UV-spectrophotometric estimation of teneligliptin from tablet dosage form. As J Pharm Anal Med Chem. 2016;4:148-156.
  13. Sowjanya P. RP-HPLC method development of metformin in pharmaceutical dosage form research and reviews. J Pharm Anal. 2015;4:2347-2340.
  14. Madhukar A, Prince A, Vijay Kumar R, Sanjeeva Y, Jagadeeshwar K, Raghupratap D. Simple and sensitive analytical method development and validation of Metformin Hydrochloride by RP-HPLC. Int J Pharm Pharml Sci. 2011;3:117-120.
  15. AliSaad AA, Mukkanti K, Chandrashekar K. Analytical method development and method validation of metformin hydrochloride and pioglitazone hydrochloride in bulk and their pharmaceutical dosage forms by RP-HPLC. Anal Chem Ind J. 2011;10:771-775.
  16. Murthy TGK, Geethanjali J. Development of a validated RP-HPLC method for simultaneous estimation of metformin hydrochloride and rosuvastatin calcium in bulk and in-house formulation. Chrom Sep Tech. 2014;5:1-7.
  17. Neelima K, Prasad RY. Analytical method development and validation of metformin, voglibose, glimepiride in bulk and combined tablet dosage form by gradient RP-HPLC. Pharm Met. 2014;5:27-33.
  18. Prasad PB, Satyanaryana K, Krishnamohan G. Development and validation of a method for simultaneous determination of metformin and saxagliptin in a formulation by RP-HPLC. Am J Anal Chem. 2015;6:841-850.
  19. Shailesh VL, Kamna RP, Jani GK, Sachin BN. Simultaneous estimation of teneligliptin hydrobromide hydrate and its degradation product by RP-HPLC method. J Pharm Sci Biosci Res. 2016;6:254-261.
  20. Shinde VC, Aher KB, Bhavar GB, Kakad SJ, Chaudhari SR. Development and validation of UV spectrophotometric method and high performance thin layer chromatographic (HPTLC) method for estimation of teneligliptin hydrobromide in pharmaceutical preparation. Der Pharm Lett. 2016;8:291-301.
  21. Chunduri RHB, Dannana GS. Development and validation of LC-MS/MS method for quantification of teneligliptin in human plasma and its application to a pharmacokinetic study. W J Pharm Pharm Sci. 2016;5:833-850.
  22. Sen AK, Hinsu DN, Sen DB, Zanwar AS, Maheshwari RA, Chandrakar VR. Analytical method development and validation for simultaneous estimation of teneligliptin hydrobromide hydrate and metformin hydrochloride from its pharmaceutical dosage form by three different UV spectrophotometric methods. J App Pharm Sci. 2016;6:157-165.
  23. Joshi H, Khristi A. Absorbance ratio method development and validation for the simultaneous estimation of teneligliptin hydrobromide hydrate and metformin hydrochloride in tablet dosage form. Int Res J Pharm. 2018;9:47-55.
  24. Irache GS, Bhajipale NS, Gandhi LR. RP-HPLC method development and validation of teneligliptin and metformin in pharmaceutical dosage forms. Int Res J Pharm. 2017;8:52-55.
  25. Deepak P, Sufiyan A, Shastry VM, Tabrej M, Lalit T. Analytical method development and validation for the simultaneous estimation of metformin and teneligliptin by RP-HPLC in bulk and tablet dosage forms. J Pharm Res. 2017;11:676-681.
  26. Manish DP, Mayank B, Priyanka S, Suleman SK. Development and validation of analytical method for simultaneous estimation of metformin hydrochloride and teneligliptin hydrobromide hydrate in pharmaceutical dosage form. J Pharm Sci Bioscientific Res. 2017;7:200-208.
  27. Shaikh AR, Ahmed BARK, Ibrahim M. A validated stability indicating RP - HPLC method for simultaneous estimation of metformin and teneligliptin in bulk and pharmaceutical dosage form. Int J Pharm Sci Res. 2018;9:1705-1712.
  28. Mrunal CS, Annapurna MM. Simultaneous determination of metformin and teneligliptin by liquid chromatography in tablets. Asian J Pharmaceu. 2018;12:736-739.
  29. ICH Harmonized Tripartite Guideline: Text on Validation of Analytical Procedures, Text and Methodology, Q2(R1), International Conference on Harmonization, Geneva; 2005:1-17.
  30. ICH Harmonized Tripartite Guideline: Stability Testing of New Drug Substances and Products, Q1A(R2), International Conference on Harmonization, Geneva; 2003:1-24.